Dart SVP with source and ipod
  Components of DART®

 

  What is controlled

  • High Voltage
  •   Grid -530 to 530 volts
      Needle -5,000 volts

  • Gas Flow
  •   Switch between Helium and Nitrogen
      Must flow when heater is running

  • Dual Automated Sample Postioners

      Send a pulse and Direction
      Integrated with medthods
      Home and Limit switches for feedback

  • Heater
  •   24v Heater
      PWM output

      PID feedback control

      Up to 550 degrees celcius